Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Más filtros










Base de datos
Intervalo de año de publicación
1.
Prehosp Emerg Care ; 24(1): 85-89, 2020.
Artículo en Inglés | MEDLINE | ID: mdl-30905240

RESUMEN

We present the case of a 30-year-old man with a subcutaneous implantable cardioverter defibrillator (S-ICD) who suffered sudden out-of-hospital cardiac arrest. During resuscitation, the patient received inappropriate shocks due to oversensing by the S-ICD of chest compression induced artifact. The rhythm on the EMS monitor showed asystole throughout the period of resuscitation. Placement of a magnet over the S-ICD generator failed to inhibit the delivery of S-ICD shocks. Information regarding inappropriate S-ICD shocks may be useful during resuscitation of patients with sudden cardiac arrest.


Asunto(s)
Reanimación Cardiopulmonar/efectos adversos , Desfibriladores Implantables/efectos adversos , Servicios Médicos de Urgencia , Insuficiencia Cardíaca/terapia , Paro Cardíaco Extrahospitalario/diagnóstico , Paro Cardíaco Extrahospitalario/terapia , Adulto , Artefactos , Resultado Fatal , Insuficiencia Cardíaca/complicaciones , Humanos , Masculino , Resultado del Tratamiento
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA
...